Awarded Contractocds-h6vhtk-03d370

High Mass Resolving Power, High Mass Accuracy, Benchtop Mass Spectrometry Instrument

The instrument must provide 480,000 mass resolving power, < 1ppm mass accuracy, a dynamic range of > 5000:1, scan rates up to 40 Hz and rapid polarity switching < 1 second per cycle. It must be supplied with an electrospray ion source.

Buyer

NATIONAL PHYSICAL LABORATORY LIMITED

Value

£800,000

Published

2 Jun 2023

Closes

2 Jun 2023

Procurement details

Source

Find a Tender

Region

United Kingdom

Procurement method

Limited

Category

Goods

Contract starts

Not provided

Contract ends

Not provided

CPV classifications

38000000

Laboratory, optical and precision equipments (excl. glasses)

Lots

Lot 1 High Mass Resolving Power, High Mass Accuracy, Benchtop Mass Spectrometry Instrument

£800,000

Our large-scale instrument development mass spectrometry imaging (MSI) studies require high mass accuracy (<1 ppm), high mass resolving power (> 400,000) and the ability to couple to a range of third party and home-built atmospheric pressure developmental ion sources. These home-built ion sources were developed for coupling to a Thermo Orbitrap Elite instrum...