Supply and Installation of an Advanced X-Ray Tomography Unit
X-Ray Tomography Device o The X-Ray source has to have an acceleration voltage of 190 kV and a beam current of 1mA o The device must have a spatial resolution of 600nm and a minimum pixel size of 100nm o The device must be supplied with different stages to test samples at different temperature ranges and under different tension/compression conditions. o The device must function fully over a temperature range from -20 C to 80 C o The devices must be able to precisely control the heating with an accuracy higher than...
Buyer
University of Birmingham
Value
Not provided
Published
20 Dec 2016
Closes
23 Jan 2017
Procurement details
Source
Contracts Finder
Region
United Kingdom
Procurement method
Open
Category
Goods
Contract starts
Not provided
Contract ends
Not provided
CPV classifications
38000000
Laboratory, optical and precision equipments (excl. glasses)
Lots
Lot default Supply and Installation of an Advanced X-Ray Tomography Unit
Value not provided
X-Ray Tomography Device o The X-Ray source has to have an acceleration voltage of 190 kV and a beam current of 1mA o The device must have a spatial resolution of 600nm and a minimum pixel size of 100nm o The device must be supplied with different stages to test samples at different temperature ranges and under different tension/compression conditions. o The...
Delivery: British Oversea Territories, Channel Islands, Europe, Isle of Man, Rest of the World, United Kingdom
Documents
Unlock document linkOpportunity notice on Contracts Finder
20 Dec 2016
Opportunity notice on Contracts Finder
Tender notice · Web page
Contains public sector information licensed under the Open Government Licence v3.0.